Agilent B1500A/B1505A Programming Guide, Edition 11 2-13
Remote Mode Functions
Measurement Modes
NOTE The PT command sets the hold time, pulse width, and pulse period. The WM
command sets the automatic abort function. The PWV sets the pulsed sweep
voltage, and the PWI sets the pulsed sweep current. The base, start, and stop values
must have the same polarity for current pulse or log sweep.
To Use
Synchronous
Sweep Source
One more channel can be set up as a staircase sweep source that has the output
synchronized with the pulsed sweep. See “Synchronous Output” on page 2-44. After
the measurement, the synchronous sweep source forces the start or stop value, as
specified by the WM command, and keeps it.
Figure 2-8 Synchronous Sweep
NOTE The WSV command sets the sweep voltage, and the WSI command sets the sweep
current. You can use the same output mode (voltage or current) as the pulsed sweep.
The start and stop values must have the same polarity for log sweep.
To Stop Sweep
Output
An automatic abort function is available. Refer to “Automatic Abort Function” on
page 2-46.
Even if the automatic abort function is disabled, the B1500 automatically stops
measurement if power compliance is enabled for the sweep source and the power
compliance or an automatic abort condition is detected.
Voltage or current
Start value
Time
Stop value
Pulse base value
Stop value
Synchronous sweep
Pulsed sweep
Previous value
Trigger (e.g. XE)
PT/WM/PWV/PWI
WSV/WSI
Start value
Previous value
Voltage or current