Agilent B1500A/B1505A Programming Guide, Edition 11 4-163
Command Reference
QSV
linteg : Integration time for the leakage current measurement, in seconds.
Numeric expression. The available values are 0.02 to 2 s for a 50 Hz
line frequency, and 0.016667 to 1.6667 s for 60 Hz. But the value is
rounded as follows:
linteg = n / selected line frequency (n : integer. 1 to 100.)
The initial setting is 5/ selected line frequency. So this value is 0.1 s for
a 50 Hz line frequency, and approx. 0.083 s for 60 Hz.
hold : Hold time (in seconds). Numeric expression. This is the time from the
start of the first sweep step to the beginning of the delay time (delay1).
0 to 655.35 sec. 0.01 sec resolution. Initial setting = 0.
delay1 : Delay time (in seconds). Numeric expression. This is the time from the
start of each sweep step to the start of the measurement.
0 to 65.535 sec. 0.0001 sec resolution. Initial setting = 0.
delay2 : Delay time (in seconds). Numeric expression. This is the time from the
end of the each sweep step to the start of the next sweep step or the end
of the sweep. delay2 is not effective for the spot measurement.
0 to 65.535 sec. 0.0001 sec resolution. Initial setting = 0.
Example
Statements
OUTPUT @B1500;"QST 0.35,0.1,5,0.2,0.2"
This example sets a hold time of 5 s and a delay time of 0.2 s for delay1 and delay2.
The integration time for the capacitance measurement is 0.35 s for a line frequency
of 60 Hz, and 0.36 s for 50 Hz. The integration time for the leakage current is 0.1 s
for 50 Hz and 60 Hz.
Where, 0.35 is equal to 21 / 60 and 17.5 / 50. This means 0.35 s is not a suitable
value for the integration time at 50 Hz. Because n must be integer. In this case, the
integration time is automatically set to 0.36 s (=18 / 50). n is rounded to 18, not 17.
QSV
This command specifies the voltage sweep source and its parameters for the
quasi-static CV measurement (MM13).
Execution
Conditions
If you do not specify the channel number for the MM command, the channel
specified by the QSV command forces the sweep voltage, and measures the
capacitance.
Syntax QSV chnum,mode,vrange,start,stop,cvoltage,step[,Icomp]
Przeglądanie stron 478