Agilent B1500A/B1505A Programming Guide, Edition 11 2-87
Remote Mode Functions
Initial Settings
Initial Settings
Agilent B1500 is initialized by turning the B1500 on, the *RST command, or the
device clear. Initial settings of the B1500 are shown in the following tables.
Table 2-9 Mainframe Settings
Setup Item Initial Setting Commands
Auto calibration off CM
Trigger mode XE, TV, TI, or GET TM
Trigger port Ext Trig In Start Measurement
trigger input
TGP
Ext Trig Out Measurement
Completion trigger
output
TGP
Digital I/O cleared TGP
Trigger delay time 0 s WT, PT
Trigger condition of
Start Step Output Setup trigger
with trigger for first sweep step TGSI
Type of output trigger edge trigger TGXO, TGSO, TGMO
Digital I/O port output for all port ERM
Program memory
cleared
a
a. Program memory is not cleared by the *RST command or the device clear.
SCR
Value of internal variable (%In, %Rn) 0 VAR
Data output format ASCII with header, CR/LF^EOI FMT
Data output buffer cleared BC
Status byte Only bit 6 is enabled. *SRE
Error code register cleared ERRX?, ERR?