4-48 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
BGV
target : Search target current (in A). Numeric expression.
0 to ±0.1 A (MPSMU/HRSMU/MCSMU)
0 to ±1 A (HPSMU/HCSMU)
0 to ±2 A (DHCSMU)
0 to ±0.008 A (HVSMU)
Remarks In the limit search mode, if search cannot find the search target and the following
two conditions are satisfied, the B1500 repeats the binary search between the last
source value and the source start value.
target is between the data at source start value and the last measurement data.
target is between the data at source stop value and the data at:
source value = | stop start | / 2.
If the search cannot find the search target and the following two conditions are
satisfied, the B1500 repeats the binary search between the last source value and the
source stop value.
target is between the data at source stop value and the last measurement data.
target is between the data at source start value and the data at:
source value = | stop start | / 2.
Example
Statements
OUTPUT @B1500;"BGI 1,0,1E-8,14,1E-6"
See Also “BSM”
BGV
The BGV command specifies the voltage monitor channel and its search parameters
for the binary search measurement (MM15). This command setting clears, and is
cleared by, the BGI command setting.
This command ignores the RV command setting.
Syntax BGV chnum,mode,condition,range,target
Parameters chnum : SMU search monitor channel number. Integer expression. 1 to 10 or
101 to 1001. See Table 4-1 on page 4-16.