Agilent B1500A/B1505A Programming Guide, Edition 11 4-229
Command Reference
WSX
Pcomp : Power compliance (in W). Numeric expression. Resolution: 0.001 W. If
the Pcomp value is not entered, the power compliance is not set. This
parameter is not available for HVSMU.
0.001 to 2 for MPSMU/HRSMU, 0.001 to 20 for HPSMU, 0.001 to 40
for HCSMU, 0.001 to 80 for dual HCSMU, 0.001 to 3 for MCSMU,
0.001 to 100 for UHVU
Example
Statements
OUTPUT @B1500;"WSV 1,0,1,100,0.01,1"
OUTPUT @B1500;"WSV 2,12,0,10"
WSX
The WSX command causes the B1500 to enter a wait state until the B1500 receives
an external trigger from a trigger input terminal specified by the port parameter. To
set the trigger logic (initial setting: negative), send the TGP command for the
specified terminal. To end a wait state before the trigger, execute the AB or *RST
command.
Syntax WSX port[,mode]
Parameters port : External trigger input port number. Integer expression. -1, or 1 to 16.
-1: Ext Trig In terminal.
1 to 16: Port 1 to 16 of the digital I/O terminal.
To use a digital I/O port, send the TGP command. The port value must
be same as the port value set to the TGP command.
mode : Waiting mode. Integer expression. 1 or 2. If this parameter is not
specified, mode is set to 1.
1: Continues the operation if an external trigger was already received.
Otherwise, the B1500 immediately goes into a wait state for an external
trigger.
2: In any condition, the B1500 immediately goes into a wait state for an
external trigger.
Remarks The B1500 checks its trigger flag to confirm the present trigger status, received or
none. To clear the trigger flag:
• Enter the *RST or device clear command (HP BASIC CLEAR statement).
• Enter the TM3 command.