4-46 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
BDT
Example
Statements
OUTPUT @B1500;"BDM 0,1"
BDT
The BDT command specifies the hold time and delay time for the quasi-pulsed
measurements.
Syntax BDT hold,delay
Parameters hold : Hold time (in sec). Numeric expression.
0 to 655.35 s, 0.01 s resolution. Initial setting is 0.
delay : Delay time (in sec). Numeric expression.
0 to 6.5535 s, 0.0001 s resolution. Initial setting is 0.
Example
Statements
OUTPUT @B1500;"BDT 0.1,1E-3"
BDV
The BDV command specifies the quasi-pulsed voltage source and its parameters.
If the output voltage is greater than the allowable voltage for the interlock open
condition, the interlock circuit must be shorted.
Syntax BDV chnum,range,start,stop[,Icomp]
Parameters chnum : SMU source channel number. Integer expression. 1 to 10 or 101 to
1001. See Table 4-1 on page 4-16.
range : Ranging type for quasi-pulsed source. Integer expression. The output
range will be set to the minimum range that covers both start and stop
values. For the limited auto ranging, the instrument never uses the
range less than the specified range. See Table 4-4 on page 4-21.
start, stop : Start or stop voltage (in V). Numeric expression. See Table 4-7 on page
4-24.
0 to ± 100 for MPSMU/HRSMU, or 0 to ± 200 for HPSMU
|start - stop| must be 10 V or more.
Icomp : Current compliance (in A). Numeric expression. See Table 4-7 on page
4-24.
If you do not set Icomp, the previous value is used.