Agilent B1500A/B1505A Programming Guide, Edition 11 2-69
Remote Mode Functions
HVSMU Current Expander
HVSMU Current Expander
The Agilent N1266A is a current expander for HVSMU. The N1266A is used to
configure the high voltage medium current unit (HVMCU) with the HVSMU
module and two MC/HCSMU modules as shown in Figure 2-42. Selector is initially
installed for switching the HVSMU or the HVMCU connected to the DUT.
To configure HVMCU, use the following command.
• ERHVCA: Specifies the modules connected to the V Control, I Control, and
HVSMU inputs.
To control the expander input-to-output path, use the following commands.
• ERHVP: Set the connection status of the expander input-to-output path.
• ERHVS: Set the connection status of the series resistor on the HVSMU path.
Figure 2-42 To Configure HVMCU
N1266A HVSMU current expander
HVSMU
GNDU
V-control
I-control
Digital I/O Digital I/O
MC/HCSMU
MC/HCSMU
B1505A
HVSMU
GNDU
Low
DUT
Drain
Source
High
HVMCU
*1: N1261A-002/004 protection adapter is required for connecting prober.
Not requred for connecting N1258A, N1259A, N1259A-300, or N1265A.
N1261A-002
or 004(*1)
Open