Agilent B1500A/B1505A Programming Guide, Edition 11 4-133
Command Reference
MCPWNX
measurement resource type. For the log sweep or using 3000 V
range of HVSMU, base, start, and stop must have the same polarity.
For using UHVU, base and peak must have the same polarity.
For current source (mode=2): See Table 4-6 on page 4-23, Table 4-8
on page 4-25, Table 4-11 on page 4-27, or Table 4-13 on page 4-28
for each measurement resource type. base, start, and stop must have
the same polarity.
comp : Compliance (in A or V). Numeric expression. If you do not set comp,
the previous value is used.
For voltage source (mode=1): See Table 4-7 on page 4-24, Table
4-9 on page 4-26, Table 4-10 on page 4-26, Table 4-12 on page
4-27, Table 4-14 on page 4-28, or Table 4-15 on page 4-28 for each
measurement resource type.
For current source (mode=2): See Table 4-6 on page 4-23, Table 4-8
on page 4-25, Table 4-11 on page 4-27, or Table 4-13 on page 4-28
for each measurement resource type.
For the log sweep and without Pcomp, set the value available for
the minimum range that covers base, start, and stop values.
Compliance polarity is automatically set to the same polarity as the
output value, regardless of the specified comp. If the output value is 0,
the polarity is set to positive.
If you set Pcomp, the maximum comp value for the measurement
resource is allowed, regardless of the output range setting.
Pcomp : Power compliance (in W). Numeric expression. Resolution: 0.001 W. If
the Pcomp value is not entered, the power compliance is not set. The
power compliance operation is based on the large one either pulse peak
or base. This parameter is not available for HVSMU.
0.001 to 2 for MPSMU/HRSMU, 0.001 to 20 for HPSMU, 0.001 to 400
for HCSMU, 0.001 to 800 for dual HCSMU, 0.001 to 30 for MCSMU,
0.001 to 22500 for UHCU, 0.001 to 200 for UHVU, 0.001 to 600 for
HVMCU
Remarks The N value and the chnum value set to the MCPNX, MCPWNX, and WNX
commands must be unique for each command execution. If you set the value used to
the previous command, the previous command setting is cleared, and the last
command setting is effective.