1-32 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Basics
Data Output Format
F Channel number of the source/measurement module. One character. See Table 1-11
on page 1-34.
G Data type. One character. Also see Table 1-12 on page 1-35.
Table 1-7 Source Data Status
Table 1-8 SMU Status
Table 1-9 CMU Status
A or EEE Explanation
W Data is for the first or intermediate sweep step.
E Data is for the last sweep step.
EEE Explanation
1 A/D converter overflowed.
2 Oscillation or force saturation occurred.
4 Another unit reached its compliance setting.
8 This unit reached its compliance setting.
16 Target value was not found within the search range.
32 Search measurement was automatically stopped.
64 Invalid data is returned. D is not used.
128 EOD (End of Data).
EEE Explanation
1 A/D converter overflowed.
2 CMU is in the NULL loop unbalance condition.
4 CMU is in the IV amplifier saturation condition.
8 Not assigned.
16 Not assigned.
32 Not assigned.
64 Invalid data is returned. D is not used.
128 EOD (End of Data).