Agilent B1500A/B1505A Programming Guide, Edition 11 4-137
Command Reference
MM
MM mode,chnum[,chnum[,chnum...[,chnum]...]]
A maximum of ten channels can be set. For mode=18, the first chnum must be
MFCMU.
• mode= 3, 4, 5, 17, 19, 20, 22, 23, or 26:
MM mode,chnum
• mode= 9 or 13:
MM mode[,chnum]
• mode= 14 or 15:
MM mode
Parameters mode : Measurement mode. Integer expression. 1 to 28. See Table 4-28.
chnum: Measurement channel number. Integer expression. See Table 4-1 on
page 4-16.
Remarks The SMU operation mode is defined by the CMM command.
The measurement range is defined by the RI or RV command (for SMU), or the RC
command (for MFCMU).
To execute the measurement, enter the XE command.
For mode=1, 2, or 10, if you use multiple measurement channels, the channels start
measurement in the order defined in the MM command.
For mode=16, if you use multiple measurement channels, the channels that use the
high speed ADC with the fixed ranging mode start measurement simultaneously,
then other channels start measurement in the order defined in the MM command.
For mode=9 or 13, if you do not specify chnum, the B1500 uses the channel
specified by the BDV or QSV command to execute measurement.
For mode=9, 14, or 15, the time stamp function is not available. See “Data Output
Format” on page 1-25.
For mode=18, MFCMU must be specified for the first chnum. Also SMU can be
used for the additional measurement channels. Then the SMU will execute the
measurement simultaneously or in the order specified by the MM command, and
finally the MFCMU will execute the measurement. The chnum parameter must not
specify the SMU connected to the SCUU.
Example
Statements
OUTPUT @B1500;"MM 1,1"
OUTPUT @B1500;"MM 2,1,3"