Agilent B1500A/B1505A Programming Guide, Edition 11 2-15
Remote Mode Functions
Measurement Modes
6. The B1500 repeats 4 and 5 for all sweep steps.
7. After the sweep measurement, the sweep sources force the start or stop value, as
specified by the WM command, and keep it.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
NOTE The WT command sets the hold time, delay time, and step delay time. The WM
command sets the automatic abort function and the output after measurement. The
WV command sets the sweep voltage, and the WI command sets the sweep current.
The WNX command sets the synchronous sweep output. The start and stop values
must have the same polarity for log sweep.
To Stop Sweep
Output
An automatic abort function is available. Refer to “Automatic Abort Function” on
page 2-46.
Even if the automatic abort function is disabled, the B1500 automatically stops
measurement if power compliance is enabled for the sweep source and the power
compliance or an automatic abort condition is detected.