5-24 Agilent B1500A/B1505A Programming Guide, Edition 11
Error Messages
Operation Error
3320 Measurement delay is out of absolute limits.
Check the measurement delay and set the correct value. The value must
be -50 ns to 50 ns, in 625 ps resolution.
3321 VM/IM measurement mode is invalid.
Check the measurement mode and set the correct value.
3322 Voltage measurement range is invalid.
Check the voltage measurement range and set the correct value.
3323 Current measurement range is invalid.
Check the current measurement range and set the correct value.
3324 WGMA?,WGMB? command query size is out of absolute limits.
Check the data size for WGMA? or WGMB? and set the correct value.
3325 Specified count for spot measurement is out of absolute limits.
Check the count value for WGMS? and set the correct value.
3326 Specified interval for spot measurement is out of absolute limits.
Check the interval value for WGMS? and set the correct value.
3327 Specified operation mode is invalid for spot measurement.
Operation mode must be DC to perform spot measurement.
4304 HV/HC/HP SMU selector must be open state.
Set the module selector input-output path to the open status to execute
the specified command.
4305 External relay control is not active.
Enable the external relay control function by using the ERHPE
command to enter the specified command. Also, check the cable
connection and the relay operation.
NOTE If one of the errors 4401 to 4408 occurs, the all module output is changed to 0 V and
the all output switch is disconnected.
4401 HVSMU over voltage emergency occurred.
4402 HVSMU over current or guard abuse emergency occurred.
4403 HVSMU guard abuse emergency occurred.