5-20 Agilent B1500A/B1505A Programming Guide, Edition 11
Error Messages
Operation Error
2122 Specified ODSW delay timing parameter out of absolute limits.
Set the appropriate delay value to ODSW.
2123 Specified ODSW width timing parameter out of absolute limits.
Set the appropriate width value to ODSW.
2131 Delay + Interval * N must be within Period value (ADC Timing).
Set the appropriate value to CORRSER?. The period value must be
more than delay+interval×count value.
2132 Specified delay for DUT impedance measurement out of absolute
limits.
Set the appropriate delay time value to CORRSER?.
2133 Specified interval for DUT impedance measurement out of absolute
limits.
Set the appropriate interval value to CORRSER?.
2134 Specified count for DUT impedance measurement out of absolute
limits.
Set the appropriate count value to CORRSER?.
2151 ALWG Sequence Data is not ready.
Sequence data must be set by using ALS before starting the output.
2152 Specified ALWG Sequence Data size is out of absolute limits.
Set the appropriate sequence data to ALS. Too large data was specified.
2153 Specified pattern index of ALWG Sequence Data is out of absolute
limits.
Set the appropriate sequence data to ALS. The pattern index in the
sequence data must be the index of a pattern defined in the pattern data.
2154 Specified repeat count of ALWG Sequence Data is out of absolute
limits.
Set the appropriate sequence data to ALS. The repeat count in the
sequence data must be 1 to 1048576.
2155 ALWG Pattern Data is not ready.
Pattern data must be set by using ALW before starting the output.