Agilent Technologies B1500A Instrukcja Obsługi Strona 542

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 598
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 541
4-226 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
WSI
2: In any condition, the B1500 immediately goes into a wait state for an
external trigger.
Remarks The B1500 checks its trigger flag to confirm the present trigger status, received or
none. To clear the trigger flag:
Enter the *RST or device clear command (HP BASIC CLEAR statement).
Enter the TM3 command.
Enter the TM1, TM2, or TM4 command to change the mode from TM3.
Enter the OS command.
Trigger the B1500 to start measurement via the Ext Trig In terminal.
Trigger the B1500 to recover from wait state set by the WS command via the
Ext Trig In terminal.
Example
Statements
OUTPUT @B1500;"WS 2"
WSI
The WSI command specifies the staircase sweep current source (synchronous sweep
source) that will be synchronized with the staircase sweep current source (primary
sweep source) set by the WI command, or the pulsed sweep current source (primary
sweep source) set by the PWI command.
Execution
Conditions
Available for the staircase sweep (MM 2), pulsed sweep (MM 4), or staircase sweep
with pulsed bias (MM5) measurement.
This command must be entered after the WI or PWI command that clears the WSI
command setting. The WV and PWV command also clears the WSI setting.
Syntax WSI chnum,range,start,stop[,Vcomp[,Pcomp]]
Parameters chnum : SMU synchronous sweep source channel number. Integer expression. 1
to 10 or 101 to 1001. See Table 4-1 on page 4-16.
range : Ranging type for synchronous sweep current output. Integer
expression. See Table 4-5 on page 4-22.
For the linear sweep, the B1500 uses the minimum range that covers
both start and stop values to force the staircase sweep current.
For the log sweep, the B1500 uses the minimum range that covers the
output value, and changes the output range dynamically.
Przeglądanie stron 541