1-28 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Basics
Data Output Format
Staircase sweep,
Multi channel
sweep,
Multi channel
pulsed sweep,
CV (DC bias)
sweep
Block1 [,Block2] . . . . <terminator>
Block1 is the block of data measured at the first sweep point.
Block2 is the block of data measured at the second sweep point.
where Block consists of the following data:
Data1 [,Data2] . . . . [,Source_data]
DataN (N: integer) is the value measured by a channel. The order of Data is defined
by the MM command. Source_data is the sweep source output value. It is sent if the
data output is enabled by the FMT command.
For the CMU measurement data, DataN will be Para1,Para2 [,Osc_level,Dc_bias].
Para1 and Para2 are the primary and secondary parameters (ex: Cp and G). They
are selected by the IMP command. See Table 4-16 on page 4-29. And Osc_level and
Dc_bias are the monitor values of the oscillator level (AC signal level) and the DC
bias respectively. They are sent if the data output is enabled by the LMN command.
Pulsed sweep,
Staircase sweep
with pulsed bias,
Pulsed sweep CV,
CV (AC level)
sweep,
C-f sweep
Block1 [,Block2] . . . . <terminator>
Block1 is the block of data measured at the first sweep point.
Block2 is the block of data measured at the second sweep point.
where Block consists of the following data:
Data [,Source_data]
Data is the value measured by the channel you specify by using the MM command.
Source_data is the sweep source output value. It is sent if the data output is enabled
by the FMT command.
For the pulsed sweep CV measurement, Data will be Para1,Para2.
For the CV (AC level) and C-f, Data will be Para1,Para2 [,Osc_level,Dc_bias].
Para1 and
Para2 are the primary and secondary parameters (ex: Cp and G). They
are selected by the IMP command. See Table 4-16 on page 4-29. And Osc_level and
Dc_bias are the monitor values of the oscillator level (AC signal level) and the DC
bias respectively. They are sent if the data output is enabled by the LMN command.
Komentarze do niniejszej Instrukcji