4-60 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
CLCORR
CLCORR
This command disables the MFCMU open/short/load correction function and clears
the frequency list for the correction data measurement. This command also clears
the correction data.
Syntax CLCORR chnum,mode
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
mode : Command option. Integer expression. 1 or 2.
1: Just clears the frequency list.
2: Clears the frequency list and sets the default frequencies, 1 k, 2 k,
5 k, 10 k, 20 k, 50 k, 100 k, 200 k, 500 k, 1 M, 1.2 M, 1.5 M, 2 M,
2.5 M, 2.7 M, 3 M, 3.2 M, 3.5 M, 3.7 M, 4 M, 4.2 M, 4.5 M, 5 MHz.
Example
Statements
OUTPUT @B1500;"CLCORR 9,1"
CM
The CM command sets the SMU auto-calibration function to ON or OFF. If the
following two conditions are satisfied, the B1500 automatically calibrates all
modules every 30 minutes.
Auto-calibration is ON.
For all SMUs, the output switch has been OFF for 30 minutes.
Syntax CM mode
Parameters mode : Auto-calibration ON or OFF. Integer expression.
0: OFF (initial setting)
1: ON
Remarks Before starting the calibration, open the measurement terminals.
If the auto-calibration is enabled, do not forget to open the measurement terminals
after measurements.
Example
Statements
OUTPUT @B1500;"CM 0"
OUTPUT @B1500;"CM 1"