3-36 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Examples
Quasi Pulsed Spot Measurements
Quasi Pulsed Spot Measurements
To perform quasi-pulsed spot measurements, use the following commands.
Function Command Parameters
Enables channels CN [chnum ... [,chnum] ... ]
Disables channels CL [chnum ... [,chnum] ... ]
Sets filter ON/OFF [FL] mode[,chnum ... [,chnum] ... ]
Sets series resistor ON/OFF [SSR] chnum,mode
Sets integration time
(Agilent B1500 can use
AAD/AIT instead of AV.)
[AV] number[,mode]
[AAD] chnum[,type]
[AIT] type,mode[,N]
Sets detection interval [BDM] interval[,mode]
Sets timing parameters [BDT] hold,delay
Sets quasi-pulsed source BDV chnum,range,start,stop[,comp]
Forces constant voltage DV, TDV chnum,range,output
[,comp[,polarity[,crange]]]
Forces constant current DI, TDI
Sets voltage measurement range [RV] chnum,range
Sets current measurement range [RI] chnum,range
[RM] chnum,mode[,rate]
Sets measurement mode MM 9[,chnum]
Sets SMU operation mode [CMM] chnum,mode
Executes measurement XE