Agilent B1500A/B1505A Programming Guide, Edition 11 4-59
Command Reference
CL
CL
The CL command disables the specified channels.
Execution
Conditions
No channel may be in the high voltage state (forcing more than the allowable
voltage for the interlock open condition, or voltage compliance set to more than it).
However, if you do not specify chnum for CL command, there are no restrictions on
the execution conditions.
Syntax CL [chnum[,chnum...[,chnum]...]]
A maximum of 15 channels can be set.
Parameters chnum : Channel number. Integer expression. See Table 4-1 on page 4-16.
If you specify multiple chnums, the channels will be disabled in the specified order.
If you do not specify chnum, this command disables all SMU, all SPGU, and CMU
in this order. Then, SMU will be disabled in the order from higher to lower output
range and SPGU will be disabled in the order from higher to lower setup voltage.
Remarks The CL command sets the specified module to the following conditions:
After this command, there is no additional power consumption for the idle state.
Example
Statements
OUTPUT @B1500;"CL"
OUTPUT @B1500;"CL 1,2,3,5"
SMU setup parameter Value MFCMU setup parameter Value
Output switch OFF DC bias 0 V
Source mode Voltage AC level 0 V
Output voltage 0 V Output signal frequency 1 kHz
V range 20 V Measurement range 50 Ω
I compliance 100 μA SPGU setup parameter Value
I range 100 μA Output switch OFF
Filter OFF Output mode DC
Series resistor Not changed Output voltage 0 V