4-50 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
BSI
Example
Statements
OUTPUT @B1500;"BGV 1,0,0.1,12,5"
See Also “BSM”
BSI
The BSI command sets the current search source for the binary search measurement
(MM15). After search stops, the search channel forces the value specified by the
BSM command.
This command clears the BSV, BSSI, and BSSV command settings.
This command setting is cleared by the BSV command.
If Vcomp value is greater than the allowable voltage for the interlock open
condition, the interlock circuit must be shorted.
Syntax BSI chnum,range,start,stop[,Vcomp]
Parameters chnum : SMU search source channel number. Integer expression. 1 to 10 or 101
to 1001. See Table 4-1 on page 4-16.
range : Output ranging type. Integer expression. The output range will be set to
the minimum range that covers both start and stop values. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See Table 4-5 on page 4-22.
start, stop : Search start or stop current (in A). Numeric expression. See Table 4-6
on page 4-23, Table 4-8 on page 4-25, or Table 4-11 on page 4-27 for
each measurement resource type. The start and stop must have different
values.
Vcomp : Voltage compliance value (in V). Numeric expression. See Table 4-6 on
page 4-23, Table 4-8 on page 4-25, or Table 4-11 on page 4-27 for each
measurement resource type. If you do not specify Vcomp, the previous
value is set.
Example
Statements
OUTPUT @B1500;"BSI 1,0,1E-12,1E-6,10"