Agilent B1500A/B1505A Programming Guide, Edition 11 4-219
Command Reference
WMACV
If the measurement is stopped by the automatic abort function, the staircase sweep
sources force the start value, and the pulsed sweep source forces the pulse base
value after sweep.
Syntax WM abort[,post]
Parameters abort : Automatic abort function. Integer expression.
1: Disables the function. Initial setting.
2: Enables the function.
post : Source output value after the measurement is normally completed.
Integer expression.
1: Start value. Initial setting.
2: Stop value.
If this parameter is not set, the sweep sources force the start value.
Output Data The B1500 returns the data measured before an abort condition is detected. Dummy
data 199.999E+99 will be returned for the data after abort.
Example
Statements
OUTPUT @B1500;"WM 2"
OUTPUT @B1500;"WM 2,2"
WMACV
This command enables or disables the automatic abort function for the CV (AC
level) sweep measurement (MM23). The automatic abort function stops the
measurement when one of the following conditions occurs.
NULL loop unbalance condition
IV amplifier saturation condition
Overflow on the AD converter
This command also sets the post measurement condition of the MFCMU. After the
measurement is normally completed, the MFCMU forces the value specified by the
post parameter.
If the measurement is stopped by the automatic abort function, the MFCMU forces
the start value.
Syntax WMACV abort[,post]