Agilent Technologies B1500A Instrukcja Obsługi Strona 476

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 598
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 475
4-160 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
QSM
Parameters data : Leakage current data output. Integer expression. 0 or 1.
0 : Disables data output. Initial setting.
1: Enables data output.
compen : Leakage current compensation. Integer expression. 0 or 1.
0 : Disables compensation. Initial setting.
1: Enables compensation.
Example
Statements
OUTPUT @B1500;"QSL 0,0"
If you send the above command, the leakage current is not measured during the
quasi-static CV measurements.
QSM
This command enables or disables the automatic abort function for the quasi-static
CV measurement (MM13). The automatic abort function stops the measurement
when one of the following conditions occurs.
Compliance on the measurement channel
Compliance on the non-measurement channel
Overflow on the AD converter
Oscillation on any channel
This command also sets the post measurement condition for the sweep source. After
the measurement is normally completed, the sweep source forces the value specified
by the post parameter.
If the measurement is stopped by the automatic abort function, the sweep source
forces the start value.
Syntax QSM abort[,post]
Parameters abort : Automatic abort function. Integer expression. 1 or 2.
1: Disables the function. Initial setting.
2: Enables the function.
post : Source output value after the measurement is normally completed.
Integer expression. 1 or 2.
1: Start value. Initial setting.
Przeglądanie stron 475
1 2 ... 471 472 473 474 475 476 477 478 479 480 481 ... 597 598

Komentarze do niniejszej Instrukcji

Brak uwag