Agilent B1500A/B1505A Programming Guide, Edition 11 4-139
Command Reference
MSC
MSC
The MSC command enables or disables the automatic abort function for the
sampling measurement (MM10 and MM26). The automatic abort function stops the
measurement when one of the following conditions occurs:
Compliance on the measurement channel (MM10)
Compliance on the non-measurement channel (MM10)
Oscillation on any channel (MM10)
NULL loop unbalance condition (MM26)
IV amplifier saturation condition (MM26)
Overflow on the AD converter (MM10 and MM26)
This command also sets the post measurement condition for the sources. After the
measurement is normally completed, the voltage/current sources force the value
specified by the post parameter. The post parameter is effective only for the DC
output channels in MM10.
If the measurement is stopped by the automatic abort function, the voltage/current
sources force the base value.
Syntax MSC abort[,post]
Parameters abort : Automatic abort function. Integer expression.
1: Disables the function. Initial setting.
2: Enables the function.
post : Source output value after the measurement is normally completed.
Integer expression.
1: Base value.
2: Bias value. Initial setting.
If this parameter is not set, the sources force the bias value.
a. Pulse current output or pulse voltage output
b. Pulse voltage output
c. DC voltage output or pulse voltage output