Agilent B1500A/B1505A Programming Guide, Edition 11 4-81
Command Reference
ERHPA?
ERHPA?
The ERHPA? command returns the channel numbers of the measurement resources
connected to the input ports of Agilent N1258A/N1259A module selector.
Syntax ERHPA?
Query Response hvsmu,hcsmu,hpsmu<CR/LF^EOI>
hvsmu : Channel number of HVSMU connected to the HVSMU port directly or
via the N1266A expander.
hcsmu : Channel number of HCSMU or DHCSMU connected to the HCSMU
port
smu : Channel number of MPSMU or HPSMU connected to the HPSMU port
Channel number will be 1 to 10, or 101 to 1001. Integer expression. Or 0 is returned
if the associated module has not been installed.
Example
Statements
OUTPUT @B1500;"ERHPA?"
ENTER @B1500;A,B,C
ERHPE
Only for the N1258A module selector users. This command enables or disables the
external relay control function. See “External Relay Control Output” on page 2-65.
Execution
Conditions
Digital I/O port must be set to the N1258A/N1259A control mode using the
ERMOD 2 command.
Syntax ERHPE onoff
Parameters onoff : 1 (enable) or 0 (disable, initial setting). Integer expression.
Example
Statements
OUTPUT @B1500;"ERHPE 1"
ERHPE?
Only for the N1258A module selector users. This command returns the condition of
the external relay control function.
Syntax ERHPE?
Query Response onoff<CR/LF^EOI>