4-186 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
SSP
Syntax SSP chnum,path
Parameters chnum : MFCMU channel number. Integer expression. 3 to 10 or 301 to 1001.
See Table 4-1 on page 4-16.
path : Path connected to the SCUU output. 1 to 4. See Table 4-30.
Example
Statements
OUTPUT @B1500;"SSP 9,4"
Remarks When the connection is changed from SMU to MFCMU, the SMU output will be set
as follows. The other setup parameters are not changed.
Output voltage 0 V
Output range 100 V
Compliance 20 mA
Series resistance OFF
When the connection is changed from MFCMU to SMU, the SMU output will be set
as follows. The other setup parameters are not changed.
Output voltage 0 V
Output range 20 V
Compliance 100 μA
Series resistance Condition before the connection is changed from SMU to
MFCMU
Table 4-30 SCUU Input Output Connection Control
Force1/Sense1 is connected to the SMU installed in the slot numbered slot-1.
Command
SCUU output connection after the command
CMUH/Force1/Sense1 CMUL/Force2/Sense2
SSP chnum, 1 Force1/Sense1 Open
SSP chnum, 2 Open Force2/Sense2
SSP chnum, 3 Force1/Sense1 Force2/Sense2
SSP chnum, 4 CMUH CMUL