Agilent Technologies B1500A Instrukcja Obsługi Strona 354

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 598
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 353
4-38 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
AIT
1: Perform the phase compensation data measurement.
If the mode parameter is not set, mode=1 is set.
Query Response results<CR/LF^EOI>
results returns the following value.
If the phase compensation measurement has never been performed, result=3 is
returned.
Example
Statements
OUTPUT @B1500;"ADJ?"
ENTER @B1500;A
AIT
This command is used to set the operation mode and the setup parameter of the A/D
converter (ADC) for each ADC type.
Execution
Conditions
Enter the AAD command to specify the ADC type for each measurement channel.
Syntax AIT type,mode[,N]
Parameters type : Type of the A/D converter. Integer expression. 0, 1, or 2.
0: High-speed ADC
1: High-resolution ADC. Not available for the HCSMU, HVSMU and
MCSMU.
2: High-speed ADC for pulsed-measurement
mode : ADC operation mode. Integer expression. 0, 1, 2, or 3.
0: Auto mode. Initial setting.
1: Manual mode
results Meaning
0 Phase compensation measurement was normally completed.
1 Phase compensation measurement failed.
2 Phase compensation measurement was aborted.
3 Phase compensation measurement has not been performed.
Przeglądanie stron 353
1 2 ... 349 350 351 352 353 354 355 356 357 358 359 ... 597 598

Komentarze do niniejszej Instrukcji

Brak uwag