Agilent B1500A/B1505A Programming Guide, Edition 11 4-179
Command Reference
SPRM?
condition : Number of pulses or sequences to output, or output duration (seconds).
Numeric expression. The following values are valid for the condition
parameter.
When mode = 1, 1 (initial value) to 1,000,000 times.
When mode = 2, IE-6 (initial setting) to 31,556,926 seconds (1 year),
setting resolution 1E-8 seconds.
Example
Statements
OUTPUT @B1500;"SPRM 1,300"
SPRM?
This query command returns the operating mode and settings of the SPGU channel
output.
Syntax SPRM?
Query Response mode[,condition]<CR/LF^EOI>
mode : Returns the operating mode of the SPGU channel output. Integer
expression.
0: Free run mode. condition is not returned.
1: Count mode.
2: Duration mode.
condition : Returns the number of pulses or sequences to output, or output duration
(seconds). Numeric expression.
When mode = 1, returns the number of outputs (times).
When mode = 2, returns the duration of the output (seconds).
Example
Statements
OUTPUT @B1500;"SPRM?"
ENTER @B1500;A,B
SPST?
This query command returns the waveform output state of the SPGU.
Syntax SPST?
Query Response state<CR/LF^EOI>
0: Pulse output inactive or ALWG sequence output inactive