Agilent Technologies B1500A Instrukcja Obsługi Strona 297

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 598
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 296
Agilent B1500A/B1505A Programming Guide, Edition 11 3-119
Programming Examples
Using Trigger Function
The following commands are also available to send a trigger or wait for an external
trigger input. Refer to “Using Trigger Function” on page 2-80.
Command Function and Syntax
OS Causes the Agilent B1500 to send a trigger signal from the Ext
Trig Out terminal.
OS
OSX
a
a. Enter the TGP command to set the trigger port.
Causes the Agilent B1500 to send a trigger signal from the
specified port.
OSX port[,level]
WS Enters a wait state until the Agilent B1500 receives an external
trigger via the Ext Trig In terminal.
WS [mode]
WSX
a
Enters a wait state until the Agilent B1500 receives an external
trigger via the specified port.
WSX port[,mode]
PA Pauses command execution or internal memory program
execution until the specified wait time has elapsed, or until a
trigger is received from the Ext Trig In terminal if the TM3
command has been entered.
PA [wait]
PAX
a
Pauses command execution or internal memory program
execution until the specified wait time has elapsed, or until a
trigger is received from the specified port if the TM3 command
has been entered.
PAX port[,wait]
TGP Sets trigger port to the specified terminal.
TGP port,terminal,polarity[,type]
TM3 Uses an external trigger to release the PA/PAX command state or
to start measurement when the B1500 is not in the PA/PAX/
WS/WSX command state.