Agilent B1500A/B1505A Programming Guide, Edition 11 3-49
Programming Examples
Multi Channel Pulsed Spot Measurements
Multi Channel Pulsed Spot Measurements
To perform multi channel pulsed spot measurements, use the following commands.
Function Command Parameters
Enables channels CN [chnum ... [,chnum] ... ]
Disables channels CL [chnum ... [,chnum] ... ]
Sets filter ON/OFF [FL] mode[,chnum ... [,chnum] ... ]
Sets series resistor ON/OFF [SSR] chnum,mode
Measurement time [AIT] 2,mode[,N]
Sets pulse common
parameters
MCPT hold[,period[,Mdelay [,average]
]]
Sets pulse timing parameters MCPNT chnum,delay,width
Sets pulse output MCPNX N,chnum,mode,range,base,pulse
[,comp]
Forces constant voltage DV, TDV chnum,range,output
[,comp[,polarity[,crange]]]
Forces constant current DI, TDI
Sets voltage measurement
range
[RV] chnum,range
Sets current measurement
range
[RI] chnum,range
[RM] chnum,mode[,rate]
Sets measurement mode MM 27,chnum[,chnum ...[,
chnum] ...]
Sets SMU operation mode [CMM] chnum,mode
Executes measurement XE