3-136 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Examples
Using Programs for 4155B/4156B/4155C/4156C
Using Programs for 4155B/4156B/4155C/4156C
This section describes the program modification example to use a FLEX command
program created for the Agilent 4155B/4156B/4155C/4156C Parameter Analyzer.
To use the program:
1. change the GPIB address, if necessary.
2. enter the ACH command to translate the channel numbers, if necessary.
3. change the FMT command parameter value to use the data output format
compatible with the 4155/4156 output data, or change the program lines to read
the measurement data.
4. remove the US command.
5. remove the RMD? command.
6. remove the unsupported command, or replace the command with the
corresponding command supported by the B1500.
For more information, refer to “To Use Programs for Agilent 4155/4156” on page
1-65.
The following program examples show a modified measurement program, which
performs a high-speed spot measurement.