Agilent B1500A/B1505A Programming Guide, Edition 11 4-97
Command Reference
ERPFTEMP?
ERPFTEMP?
This query command returns the temperature measured by using the thermocouples
(N1265A-041) connected to the K Thermocouple 1 and 2 terminals of Agilent
N1265A ultra high current expander/fixture.
Execution
Conditions
Digital I/O port must be set to the N1265A control mode using the ERMOD 4
command.
Syntax ERPFTEMP? chnum
Parameters chnum : Terminal number of the K thermocouple; 1 or 2.
Query Response value<CR/LF^EOI>
Measured temperature in degrees C. Numeric expression.
Example
Statements
OUTPUT @B1500;"ERPFTEMP? 1"
ENTER @B1500;Temp1
ERPFUHCA
This command specifies the MC/HCSMU modules connected to the V Control and I
Control inputs of Agilent N1265A ultra high current expander/fixture.
Execution
Conditions
No SMU may be in the high voltage state (forcing more than the allowable voltage
for the interlock open condition, or voltage compliance set to more than it).
The CL command has been already executed for the measurement resources
specified in the ERPFUHCA command.
Digital I/O port must be set to the N1265A control mode using the ERMOD 4
command.
Syntax ERPFUHCA vsmu,ismu
Parameters vsmu : Slot number for the MC/HCSMU connected to the V Control input.
ismu : Slot number for the MC/HCSMU connected to the I Control input.
Effective slot numbers are 1 to 10. Integer expression. Use the smaller slot number
for the HCSMU which occupies two slots.
Enter 0 to cancel the assignment.