Agilent B1500A/B1505A Programming Guide, Edition 11 4-105
Command Reference
FL
FL
This command sets the connection mode of a SMU filter for each channel.
A filter is mounted on the SMU. It assures clean source output with no spikes or
overshooting.
Syntax FL mode[,chnum[,chnum...[,chnum]...]]
A maximum of ten channels can be set.
Parameters mode : Status of the filter. Integer expression.
0: Disconnect (initial setting).
1: Connect.
chnum : SMU channel number. Integer expression. 1 to 10 or 101 to 1001. See
Table 4-1 on page 4-16.
If you do not specify chnum, the FL command sets the same mode for all channels.
Example
Statements
OUTPUT @B1500;"FL 1"
OUTPUT @B1500;"FL 0,1,3,5"
FMT
This command clears the B1500 output data buffer, and specifies the data output
format. For details about data output format, see “Data Output Format” on page
1-25. Without this command, the data output format is same as the format by the
FMT1,0 command.
Query command output data is always stored in the query buffer in ASCII format,
regardless of this command.
NOTE Multi command statement is not allowed for this command.
Syntax FMT format[,mode]
Parameters format : Data output format. Integer expression. 1 to 25. See Table 4-24.
mode : Data output mode. Integer expression. 0 to 10. See Table 4-25. Without
setting the mode value, only the measurement data is returned.
For the sampling measurement, the sampling index is also returned if
no zero value is set to mode.