Agilent B1500A/B1505A Programming Guide, Edition 11 4-65
Command Reference
CORRDT?
Example
Statements
OUTPUT @B1500;"CORRDT 9,3000000,0,0,0,0,0,0"
CORRDT?
This command returns the MFCMU open/short/load correction data.
Syntax CORRDT? chnum,index
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
index : Index number of the list. Integer expression.
Query Response freq,open_r,open_i,short_r,short_i,load_r,load_i<CR/LF^E
OI>
freq : Frequency of the correction data. Numeric expression in Hz.
open_r : Open correction data (G). Numeric expression in S.
open_i : Open correction data (B). Numeric expression in S.
short_r : Short correction data (R). Numeric expression in Ω.
short_i : Short correction data (X). Numeric expression in Ω.
load_r : Load correction data (R). Numeric expression in Ω.
load_i : Load correction data (X). Numeric expression in Ω.
Example
Statements
OUTPUT @B1500;"CORRDT 9,3000000,0,0,0,0,0,0"
OUTPUT @B1500;"CORRDT? 9,1"
ENTER @B1500;Freq,Open_r,Open_i,Short_r,Short_i,Load_r,
Load_i
CORRL
This command disables the MFCMU open/short/load correction function and adds
the specified frequency to the frequency list for the correction data measurement.
This command also clears the correction data.
Syntax CORRL chnum,freq
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
freq : Frequency to be added. Numeric expression. in Hz. See Table 4-18 on
page 4-30.