Agilent Technologies B1500A Instrukcja Obsługi Strona 529

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 598
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 528
Agilent B1500A/B1505A Programming Guide, Edition 11 4-213
Command Reference
WACV
WA CV
This command sets the AC level sweep source used for the CV (AC level) sweep
measurement (MM23). The sweep source will be the MFCMU.
Execution
Conditions
The CN/CNX command has been executed for the specified channel.
Syntax WACV chnum,mode,start,stop,step
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
mode : Sweep mode. Integer expression. 1 to 4.
1: Linear sweep (single stair, start to stop.)
2: Log sweep (single stair, start to stop.)
3: Linear sweep (double stair, start to stop to start.)
4: Log sweep (double stair, start to stop to start.)
start, stop : Start or stop value of the AC level sweep (in V). Numeric expression.
0 (initial setting) to 0.250 V, 0.001 V step.
step : Number of steps for staircase sweep. Numeric expression. 1 to 1001.
Example
Statements
OUTPUT @B1500;"WACV 9,1,0.001,0.100,100"
WAT
This command sets the source wait time and the measurement wait time as shown in
Figure 4-6. The wait time is given by the following formula:
wait time = N × initial wait time + offset
where initial wait time is the time the Agilent B1500 initially sets and you cannot
change. The initial source wait time is not same as the initial measurement wait
time. The SMU wait time settings are effective for all SMUs.