Agilent B1500A/B1505A Programming Guide, Edition 11 4-25
Command Reference
Table 4-8 MC/HC/DHCSMU Current Source Setup Parameters
1
Output
range
(actually
used)
a
a. Output ranges 20 A and 40 A are available only for pulse output, not for DC output.
Setting
resolution in
A
current, start, stop,
base, bias, or pulse
in A
Maximum Vcomp value in V
Maximum
pulse base
value
MC
SMU
HC
SMU
DHC
SMU
10 μA 1E-11 0 to ± 11.5E-6 ± 30 ± 40 ± 40 ± 10 μA
100 μA 1E-10 0 to ± 115E-6 ± 100 μA
1 mA 1E-9 0 to ± 1.15E-3 ± 1 mA
10 mA 1E-8 0 to ± 11.5E-3 ± 10 mA
100 mA 1E-7
0 to ± 115E-3
b
b. Maximum 100 mA for the DC output of MCSMU.
± 100 mA
1 A
c
c. For the MCSMU, the 1 A range is available only for the pulse output within the maximum pulse
width 100 ms and maximum duty ratio 5 %.
1E-6
0 to ± 1.15
d
d. Maximum 1 A for the MCSMU and for the DC output of HCSMU.
±1 A
e
e. Maximum 50 mA for the MCSMU.
2 A 2E-6
0 to ± 2.02
f
f. Maximum 2 A for the DC output.
NA NA ± 2 A
20 A 2E-5 0 to ± 20 ± 20 NA ± 100 mA
40 A 4E-5 0 to ± 40 NA ± 20 ± 200 mA
1. The following commands are used for setting the HCSMU, DHCSMU, and MCSMU to the cur-
rent source.
DI, TDI, WI, WSI, WNX, PI, PWI, LSI, LSSI, BSI, BSSI, MI, MCPNX, MCPWNX