Agilent Technologies B1500A Instrukcja Obsługi Strona 426

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 598
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 425
4-110 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
LGI
Syntax INTLKVTH?
Query Response voltage<CR/LF^EOI>
This value returns 0 (maximum 0 V) or 42 (maximum ±42 V).
Example
Statements
OUTPUT @B1500;"INTLKVTH?"
ENTER @B1500;Vintlk
LGI
The LGI command sets the current monitor channel for the linear search
measurement (MM14). This command setting clears, and is cleared by, the LGV
command setting.
This command ignores the RI command setting.
Syntax LGI chnum,mode,range,target
Parameters chnum : SMU search monitor channel number. Integer expression. 1 to 10 or
101 to 1001. See Table 4-1 on page 4-16.
mode : Search mode. Integer expression.
0 : If the measured value target, it is the search result data.
1 : If the measured value target, it is the search result data.
range : Measurement ranging type. Integer expression. The measurement range
will be set to the minimum range that covers the target value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See Table 4-3 on page 4-19.
target: Search target current (in A). Numeric expression.
0 to ±0.1 A (MPSMU/HRSMU/MCSMU)
0 to ±1 A (HPSMU/HCSMU)
0 to ±2 A (DHCSMU)
0 to ±0.008 A (HVSMU)
Example
Statements
OUTPUT @B1500;"LGI 0,1,14,1E-6"
LGV
The LGV command sets the voltage monitor channel for the linear search
measurement (MM14). This command setting clears, and is cleared by, the LGI
command setting.
Przeglądanie stron 425
1 2 ... 421 422 423 424 425 426 427 428 429 430 431 ... 597 598

Komentarze do niniejszej Instrukcji

Brak uwag