4-174 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
SIM
Example
Statements
OUTPUT @B1500;"SER? 101"
ENTER @B1500;A
SIM
The SIM command sets the SPGU operation mode, PG or ALWG. The setting is
effective for the all SPGU modules installed in the B1500. This command also
triggers 0 V output of the SPGU channels which output switch has been ON.
Syntax SIM mode
Parameters mode : SPGU operation mode. Integer expression. 0 or 1.
0: PG (pulse output) mode
1: ALWG (arbitrary linear wave output) mode
Example
Statements
OUTPUT @B1500;"SIM 1"
SIM?
This query command returns the present SPGU operation mode of the B1500.
Syntax SIM?
Response mode<CR/LF^EOI>
0: PG (pulse output) mode
1: ALWG (Arbitrary linear wave output) mode
Example
Statements
OUTPUT @B1500;"SIM?"
ENTER @B1500;A
SOPC
This command is effective for the HVMCU, UHCU, and UHVU. This command
enables the output power comparison function.
This function automatically calculates the power by multiplying the current by the
voltage applied to the device under test and compares the calculation result with the
power value specified in this command. Then if it is not less than the power value,
the measurement status is set to “compliance”.
This function is disabled by setting 0 to power.