Agilent B1500A/B1505A Programming Guide, Edition 11 3-9
Programming Examples
High-Speed Spot Measurements
High-Speed Spot Measurements
To perform high-speed spot measurements, use the following commands.
You can use the above commands regardless of the measurement mode (MM
command settings).
Function Command Parameters
Enables channels CN [chnum ... [,chnum] ... ]
Disables channels CL [chnum ... [,chnum] ... ]
Sets filter ON/OFF [FL] mode[,chnum ... [,chnum] ... ]
Sets series resistor ON/OFF [SSR] chnum,mode
Sets integration time
(Agilent B1500 can use
AAD/AIT instead of AV.)
[AV] number[,mode]
[AAD] chnum[,type]
[AIT] type,mode[,N]
Forces constant voltage DV, TDV chnum,range,output
[,comp[,polarity[,crange]]]
Forces constant current DI, TDI
Performs current measurement TI chnum[,range]
TTI chnum[,range]
Performs voltage measurement TV chnum[,range]
TTV chnum[,range]
Resets the time stamp TSR
Returns the time stamp at this
time
TSQ