Agilent B1500A/B1505A Programming Guide, Edition 11 2-11
Remote Mode Functions
Measurement Modes
7. After the sweep measurement, the pulsed source forces the pulse base value, and
the sweep source forces the start or stop value, as specified by the WM
command, and keeps it.
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
NOTE The WM command sets the automatic abort function and the output after
measurement. The WV command sets the sweep voltage, and the WI command sets
the sweep current. The start and stop values must have the same polarity for log
sweep.
The PT command sets the pulse timing parameters, such as pulse width and pulse
period. The PV command sets voltage pulse, and the PI command sets current pulse.
The base and peak values must have the same polarity for the current pulse.
To Use
Synchronous
Sweep Source
One more channel can be set up as a staircase sweep source that has the output
synchronized with the staircase sweep. See Figure 2-5 and “Synchronous Output
on page 2-44. After the measurement, the synchronous sweep source forces the start
or stop value, as same as the primary sweep source, and keeps it.
NOTE The WSV command sets the sweep voltage, and the WSI command sets the sweep
current. You can use the same output mode (voltage or current) as the primary
sweep. The start and stop values must have the same polarity for log sweep.
To Stop Sweep
Output
An automatic abort function is available. Refer to “Automatic Abort Function” on
page 2-46.
Even if the automatic abort function is disabled, the B1500 automatically stops
measurement if power compliance is enabled for the sweep source and the power
compliance or an automatic abort condition is detected.