Agilent B1500A/B1505A Programming Guide, Edition 11 4-55
Command Reference
BST
Icomp : Current compliance value (in A). Numeric expression. If you do not
specify Icomp, the previous value is set. Zero amps (0 A) is not a valid
value for the Icomp parameter.
Example
Statements
OUTPUT @B1500;"BSSV 1,0,5,1E-6"
See Also For the source output value, output range, and the available compliance values, see
Table 4-7 on page 4-24, Table 4-9 on page 4-26, Table 4-12 on page 4-27, or Table
4-15 on page 4-28 for each measurement resource type.
BST
The BST command sets the hold time and delay time for the binary search
measurement (MM15). If you do not enter this command, all parameters are set to 0.
Syntax BST hold,delay
Parameters hold : Hold time (in seconds) that is the wait time after starting the search
measurement and before starting the delay time for the first search
point. Numeric expression.
0 to 655.35 sec. 0.01 sec resolution.
delay : Delay time (in seconds) that is the wait time after starting to force a step
output value and before starting a step measurement. Numeric
expression.
0 to 65.535 sec. 0.0001 sec resolution.
Example
Statements
OUTPUT @B1500;"BST 5,0.1"
BSV
The BSV command sets the voltage search source for the binary search
measurement (MM15). After search stops, the search channel forces the value
specified by the BSM command.
This command clears the BSI, BSSI, and BSSV command settings.
This command setting is cleared by the BSI command.
If the output voltage is greater than the allowable voltage for the interlock open
condition, the interlock circuit must be shorted.