3-12 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Examples
Spot Measurements
Spot Measurements
To perform spot measurements, use the following commands.
NOTE If you use multiple measurement channels, the channels start measurement in the
order defined in the MM command.
Function Command Parameters
Enables channels CN [chnum ... [,chnum] ... ]
Disables channels CL [chnum ... [,chnum] ... ]
Sets filter ON/OFF [FL] mode[,chnum ... [,chnum] ... ]
Sets series resistor ON/OFF [SSR] chnum,mode
Sets integration time
(Agilent B1500 can use
AAD/AIT instead of AV.)
[AV] number[,mode]
[AAD] chnum[,type]
[AIT] type,mode[,N]
Forces constant voltage DV, TDV chnum,range,output
[,comp[,polarity[,crange]]]
Forces constant current DI, TDI
Sets voltage measurement
range
[RV] chnum,range
Sets current measurement
range
[RI] chnum,range
[RM] chnum,mode[,rate]
Sets measurement mode MM 1,chnum[,chnum ... [,chnum] ... ]
Sets SMU operation mode [CMM] chnum,mode
Executes measurement XE