Agilent B1500A/B1505A Programming Guide, Edition 11 4-87
Command Reference
ERHVP?
Digital I/O port must be set to the N1266A control mode using the ERMOD 8
command.
Syntax ERHVP state
Parameters state : Connection status of the input-to-output path. Integer expression. 0,
1,or 2.
0: Open, initial setting.
1: Connects to HVSMU
2: Connects to HVMCU
Example
Statements
OUTPUT @B1500;"ERHVP 2"
ERHVP?
This query command returns the connection status for the input-to-output path of
Agilent N1266A HVSMU current expander.
Syntax ERHVP?
Query Response state<CR/LF^EOI>
Connection status of the input-to-output path. 0, 1,or 2 (see ERHVP).
Example
Statements
OUTPUT @B1500;"ERHVP?"
ENTER @B1500;pathHVMC
ERHVPV
This command sets the operation mode of Agilent N1266A HVSMU current
expander for the performance verification.
Execution
Conditions
Digital I/O port must be set to the N1266A control mode using the ERMOD 8
command.
The ERHVP 0 command has been already executed.
The CL command has been already executed for the SMU modules specified in the
ERHVCA command.
Syntax ERHVPV state
Parameters state : Operation mode. Integer expression. 0, 1,or 2.