Agilent B1500A/B1505A Programming Guide, Edition 11 4-195
Command Reference
TGMO
TGMO
The TGMO command selects the edge trigger or the gate trigger for the Step
Measurement Completion trigger output set by the TGP port,2,polarity,3
command. See Figure 4-4.
This command is available for the staircase sweep, multi channel sweep, and
MFCMU DC/AC/frequency sweep measurements.
Figure 4-4 Trigger Output Example, Staircase Sweep Measurement, Negative Logic
Syntax TGMO mode
Parameters mode : Edge trigger or gate trigger. Integer expression. 1 or 2.
1: Edge trigger (initial setting).
2: Gate trigger.
Example
Statements
OUTPUT @B1500;"TGMO 2"
See Also “TGP” and “TGPC”
: Measurement
TGP t,2,p,3
XE
TGP t,2,p,2
TGP t,2,p,1
TGP t,2,p,3
TGP t,2,p,2
TGP t,2,p,1
Gate trigger
Edge trigger
Source trigger delay
Hold time
Delay time
Source trigger delay
Step delay time
Step Measurement Completion
Step Output Setup Completion
Measurement Completion
Step Measurement Completion
Step Output Setup Completion
Measurement Completion