Agilent Technologies B1500A Instrukcja Obsługi Strona 391

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 598
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 390
Agilent B1500A/B1505A Programming Guide, Edition 11 4-75
Command Reference
DSMPLFLUSH
Parameters chnum : SMU channel number. Integer expression. 1 to 10. See Table 4-1 on
page 4-16.
type : Event type. Integer expression.
1: Start of pulse output
count : Number of events. Integer expression. 0 disables the signal monitor.
Example
Statements
If the following command is executed, the channel 5 starts the signal monitor when
the 10th pulse output starts.
OUTPUT @B1500;"DSMPLARM 5,1,10"
Remarks The signal monitor operation is performed as a part of the MM27/MM28
measurement.
If the signal monitor operation overlaps a MM27/MM28 measurement period, the
overlapped monitor data is averaged and the result is set as a MM27/MM28
measurement data. If there is no overlap, a dummy data is set.
The MM27/MM28 measurement data is directly sent to the data output buffer
regardless of the signal monitor operation.
The channel can keep all data obtained by one execution of the signal monitor. And
the data is sent to the data buffer by the DSMPLFLUSH command.
DSMPLFLUSH
Available for HVSMU, HCSMU, MCSMU, DHCSMU, UHCU, HVMCU, and
UHVU.
The DSMPLFLUSH command sends the signal monitor data to the data output
buffer.
Execution
Conditions
The DSMPLARM command must be executed before this command.
Syntax DSMPLFLUSH chnum
Parameters chnum : SMU channel number. Integer expression. 1 to 10. See Table 4-1 on
page 4-16.
Response All data obtained by one execution of the signal monitor is returned. One data
contains the compliance side data and the force side data in this order. Time stamp
data is not included. The data output format depends on the FMT command setting.
Przeglądanie stron 390
1 2 ... 386 387 388 389 390 391 392 393 394 395 396 ... 597 598

Komentarze do niniejszej Instrukcji

Brak uwag