Agilent B1500A/B1505A Programming Guide, Edition 11 1-63
Programming Basics
Programming Tips
To Perform Quasi-Sampling Measurement
The following setup enables to perform a quasi-sampling measurement. Then the
sampling interval will be sum of delay time and step delay time.
Sets the sweep measurement mode (MM 2 or MM 16).
Sweep start value = Sweep stop value (for WI, WV, or WNX).
Sets hold time, delay time, and step delay time (WT).
To Interrupt Command Execution
The B1500 executes commands in the received order. However, only the following
commands can interrupt the command execution.
Table 1-16 Interrupt Commands
Voltage or current
Measurement trigger (e.g. XE)
Hold
Start value, Stop value
Previous value
: Measurement
Delay Step delay
Sampling interval = Delay + Step delay
Command Description
AV Changes the number of averaging samples during the
measurement.
AIT Changes the A/D converter setting of the SMU during the
measurement.
ACT Changes the A/D converter setting of the CMU during the
measurement.
AB Aborts the command execution.
*RST Resets the B1500 during the command execution.
XE If the B1500 has been set to the wait status by the PA or PAX
command, the XE command can be used to release the wait
status. For details, see Chapter 4, “Command Reference.