Agilent B1500A/B1505A Programming Guide, Edition 11 5-9
Error Messages
Operation Error
245 Specify a higher measurement range to QSR.
Too large offset current was measured. Specify the next higher
measurement range to the QSR command.
246 QSV mode value must be 1 or 3.
The mode values available for the QSV command are 1 (single linear)
and 3 (double linear). Set one of the available values.
247 Dedicated channel must be specified by QSO.
Specify the dedicated channel to the QSO command. It must be the
channel other than the measurement channel set by the MM13
command or the sweep output channel set by the QSV command.
253 Program memory is full.
Maximum of 2000 programs or 40000 commands can be stored in the
program memory. See “ST” on page 4-188.
254 Invalid input for a memory program.
The GPIB GET command (TRIGGER statement in HP BASIC) and an
external trigger input are not allowed in a memory program (between
the ST and END commands).
255 Maximum nesting level is eight.
Nesting (one program calling another) of a memory program must be
eight levels or less.
260 Data output buffer is full.
Maximum 34034 measurement data items can be stored in the data
output buffer.
270 Search source channel must be set.
Before triggering the search measurement or sending the LSSV, LSSI,
BSSV, or BSSI command to set the synchronous search source, send
the LSV, LSI, BSV, or BSI command to set the primary search source.
271 Search monitor channel must be set.
Before triggering the search measurement, send the LGV, LGI, BGV, or
BGI command to set the search monitor channel.