Agilent B1500A/B1505A Programming Guide, Edition 11 4-203
Command Reference
TMACV
Remarks In the TM3 event mode, if the B1500 is not in the wait status set by the PA, PAX,
WS, or WSX command, the B1500 can start the measurement by an external trigger
input. After measurement, the B1500 sends a trigger to a trigger output terminal. In
the initial setting, you can use the Ext Trig In and Out terminals. To use the digital
I/O port, enter the TGP command to set the trigger input or output terminal.
To set the trigger logic (initial setting: negative), send the TGP command for the
trigger input terminal.
Example
Statements
OUTPUT @B1500;"TM 1"
OUTPUT @B1500;"TM 3"
See Also “PA”, “PAX”, “TGP”, “TGPC”, “WS”, and “WSX”
TMACV
This command monitors the MFCMU AC voltage output signal level, and returns
the measurement data.
Execution
Conditions
The CN/CNX command has been executed for the specified channel.
Syntax TMACV chnum,mode[,range]
Parameters chnum : MFCMU channel number. Integer expression. 1 to 10 or 101 to 1001.
See Table 4-1 on page 4-16.
mode : Ranging mode. Integer expression. 0 or 2.
0: Auto ranging. Initial setting.
2: Fixed range.
range : Measurement range. This parameter must be set if mode=2. See Table
4-19 on page 4-30.
Example
Statements
OUTPUT @B1500;"TMACV 9,0"
ENTER @B1500 USING "#,3X,13D,X";Aclevel
TMDCV
This command monitors the MFCMU DC bias output, and returns the measurement
data.