4-20 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
range
Resource type
Ranging type
a
UHCU
b
HVMCU
0 Yes Yes Compliance range
Measurement channel uses the minimum range that covers
the compliance value.
19 Yes
21 Yes
26 Yes
28 Yes
-19 Yes 100 mA range fixed
-21 Yes 2 A range fixed
-26 Yes 500 A range fixed
-28 Yes 2000 A range fixed
a. If you specify the fixed range larger than the compliance value, the channel uses the compliance
range.
b. If the measurement resource forces current, the channel uses the current output range regardless
of the range value.
range
Resource type
Ranging type
a
UHVU DC voltage output Pulse voltage output
0 Yes Auto ranging Compliance range
Measurement channel
uses the minimum range
that covers the compliance
value.
15 Yes 10 μA limited auto
16 Yes 100 μA limited auto
17 Yes 1 mA limited auto
18 Yes 10 mA limited auto
19 Yes 100 mA limited auto
-15 Yes 10 μA range fixed
-16 Yes 100 μA range fixed
-17 Yes 1 mA range fixed
-18 Yes 10 mA range fixed
-19 Yes 100 mA range fixed
a. If you specify the fixed range larger than the compliance value, the channel uses the compliance
range.
Przeglądanie stron 335