Agilent B1500A/B1505A Programming Guide, Edition 11 4-93
Command Reference
ERPFDP?
1: Connects the Low terminal to UHCU Low, and the High terminal to
the UHCU High.
2: Connects the Low terminal to the GNDU, and the High terminal to
the HVSMU/HVMCU.
3: Connects the Low terminal to the GNDU, and the High terminal to
the MP/HPSMU.
4: Connects the Low terminal to the GNDU, and opens the High
terminal.
Example
Statements
OUTPUT @B1500;"ERPFDP 1"
ERPFDP?
This query command returns the selector input-to-output connection status of
Agilent N1265A ultra high current expander/fixture.
Syntax ERPFDP?
Query Response state<CR/LF^EOI>
Selector input-to-output connection status. 0, 1, 2, 3, or 4 (see ERPFDP).
Example
Statements
OUTPUT @B1500;"ERPFDP?"
ENTER @B1500;pathSelUHC
ERPFDS
This command sets the connection status for the series resistor on the HVSMU path
of Agilent N1265A ultra high current expander/fixture.
Execution
Conditions
Digital I/O port must be set to the N1265A control mode using the ERMOD 4
command.
The ERPFDP 0 command has been already executed.
The CL command has been already executed for the measurement resources
specified in the ERPFDA command.
Syntax ERPFDS state
Parameters state : Connection status of the series resistor on the HVSMU path. Integer
expression. 0 or 1.