82 16800 Series Portable Logic Analyzers Service Guide
5 Troubleshooting
Figure 8 Pattern Generator Troubleshooting Flowchart 1
Start
1
Apply Power
Does the display
show the module
config correctly?
Is the current
software version
installed?
Install latest operating
system version.
1A
No No
Do the procedure To run
the self tests in Chapter 3
on the suspect module.
Are ALL cards
properly seated in the
mainframe?
2x10 cable faulty or not
installed correctly
between J3 and J6 on
circuit board(s).
Check 2x10 cable(s)
Refer to Chapter 2.
Remove power and
properly seat all cards in
the mainframe.
No
Does the module
self-test pass?
Are expander boards
installed?
No
NoYes
Exit the Self Test and start
a logic analyzer session.
2
Remove power from the
mainframe.
2x10 cable faulty or not
installed correctly between
J10 and J15 or J14 and
J19 on circuit boards.
Check 2x20 cables.
Are 2x20 cables
properly seated?
Remove power, replace
defective board. Reapply
power and re-run module
self-tests.
Yes
No
Reset 2x20 cable
Yes
Yes
Yes
Yes