106 16800 Series Portable Logic Analyzers Service Guide
5 Troubleshooting
top row) are to the left of the orange transition region. If
not, see “To re- align a stray channel" on page 50.
7 Click OK to close the Thresholds and Sample Positions
dialog.
8 Click OK to close the Analyzer Setup dialog.
9 Switch to the Listing window by selecting the Listing tab
at the bottom of the main window.
10 Click the Run icon .
11 Data appears in the Listing window upon completion of
the run.
12 If the listing shows that the data alternates between
AAAA and 5555, then the probe and cable pass the test.